Can the Hach Copper methods be used to measure Copper Sulfate CuSO4?

Document ID

Document ID TE260

Version

Version 4.0

Status

Status Published

Published Date

Published Date 05/17/2018
Question
Can the Hach Copper methods be used to measure Copper Sulfate CuSO4?
Summary
Measuring copper sulfate with Hach Copper methods
Answer
Any of the copper methods can be used to measure copper sulfate.  Keep in mind that the results from the test are displayed as mg/L of copper.  Results will need to be multiplied by a conversion factor of 2.51 in order to have results as mg/L CuSO4.