Question
How do I calculate the amount of sodium thiosulfate needed to remove the chlorine in the sample?
Summary
Dechlorinating samples using sodium thiosulfate
Answer
Using solid thiosulfate: Add 7.0 mg of Product # 46001 (sodium thiosulfate, pentahydrate, ACS grade) for every milligram of chlorine.
Using the Hach thiosulfate standard solution Product # 2409353: Use 1.13 mL of 0.0250 N thiosulfate standard solution for every milligram of chlorine
The amount in milligrams of Cl2 in a sample = (concentration result from a chlorine test) x (the number of liters of sample that is being treated).
Example: 500 mL sample of 2.0 mg/L = 1 mg Cl2.
Using the Hach thiosulfate standard solution Product # 2409353: Use 1.13 mL of 0.0250 N thiosulfate standard solution for every milligram of chlorine
The amount in milligrams of Cl2 in a sample = (concentration result from a chlorine test) x (the number of liters of sample that is being treated).
Example: 500 mL sample of 2.0 mg/L = 1 mg Cl2.